RSNA 2007: Connecting Radiology
 
Home
Meeting Program
-
Subspecialty Content Brochures
-
Honors
-
Committees
Technical Exhibits
Daily Bulletin
Image Interpretation Session
Certificate of Attendance
Meeting Announcements
Attendance
R&E Foundation
RSNAnet Contributors
Career Connection
Past Meetings
Future Meetings
Newsroom
Advertising Opportunities
RSNA.org
Contact Us


TOOLS
 
  Add to Briefcase
    Print
    Email Event
PARTICIPANTS
Presenter
Jamshid Dehmeshki PhD  
Abstract Co-Author
Hamdan Amin PhD  
Anne-Marie Jouannic  
Farhang Daemi PhD  
Vahid Yaghmai MD  
Salah Qanadli  
et al  
  CODE: SST08-01
  SESSION: Vascular/Interventional (Vascular Imaging—CTA)
  Accurate Automatic Segmentation and Measurement of Aortic Aneurysm in CT Angiography
 
 
  DATE: Friday, November 30 2007
  START TIME: 10:30 AM
  END TIME: 10:40 AM
  LOCATION: E451A



  DISCLOSURES
  J.D. - Nothing to disclose.  
  H.A. - Nothing to disclose.  
  A.J. - Nothing to disclose.  
  F.D. - Nothing to disclose.  
  V.Y. - Nothing to disclose.  
  S.Q. - Nothing to disclose.  
  .e.  

 PURPOSE
 
To evaluate the performance of a new Computer-Aided Detection (CAD) system for automatic segmentation and measurement of Aortic Aneurysm (AA).
  
 METHOD AND MATERIALS
 
Fifty CT Angiography (CTA) data, from which 30 indicated abdominal AA, were used to develop and evaluate the CAD system. Three Rdaiologists manually identified and segmented all AA areas independently using in-house software. For each of the three manual segmentations, an “average” AA region was obtained by taking the overlapping part of the three manual segmentations and the result were used as our “gold standard” for development and evaluation of the automatic method. The proposed method first automatically identifies and segments the aorta lumen. Once the Aorta and its centreline was found, a series of images perpendicular to its centerline was reconstructed to measure the diameters of aorta. A method was then developed to examine all objects attached to the segmented aorta and whose characteristics lie in a mumber of pre-defined criteria set by incorporating prior understanding of the normal expected variation of aorta, such as its diameters. These regions were considered as candidates for AA regions. The identified AA regions were used to initialise a deformable model-based method used to segment and quantify Thrombus area. The accuracy of the algorithm was validated by comparing the segmentation result against the average “manual segmentation” of three radiologists. Two metrics were calculated to assess the accuracy of the algorithm: the overlapping ratio (OR) and the coverage ratio (CR)
  
 RESULTS
 
The average values for OR and CR was 0.95 (1 meaning 100% overlap), indicating that the outline segmentation of AA was very close to that defined manually by the radiologists.
  
 CONCLUSION
 
The manual identification of the enlarged portions of the aorta on a number of cross-sectional images is extremely tedious, time-consuming and subjective process. By using the Computer Aided Detection more accurate, cost effective plus faster result can be achieved.
  
 CLINICAL RELEVANCE/APPLICATION
 
The progressive growth of AA may eventually cause rupture if not diagnosed or treated.
  